Non Invasive Cardiac Output Evaluation with CO2 Rebreathing Method for CRT Patients

Authors

  • Eraldo Occhetta Cardiology Department, AOU Maggiore della Carità, Novara, Italy
  • Chiara Devecchi Cardiology Department, AOU Maggiore della Carità, Novara, Italy
  • Paolo Devecchi Cardiology Department, AOU Maggiore della Carità, Novara, Italy
  • Emanuela Facchini Cardiology Department, AOU Maggiore della Carità, Novara, Italy
  • Virginia Di Ruocco Cardiology Department, AOU Maggiore della Carità, Novara, Italy
  • Laura Plebani Cardiology Department, AOU Maggiore della Carità, Novara, Italy
  • Paolo Nicola Marino Cardiology Department, AOU Maggiore della Carità, Novara, Italy

DOI:

https://doi.org/10.6000/1929-6029.2016.05.03.3

Keywords:

Cardiac resynchronization therapy, cardiac output, CO2 rebreathing

Abstract

Background: Cardiac resynchronization therapy with ICD (CRT-D) or pacemaker (CRT-P) is useful to reverse the deleterious effects of ventricular dyssynchronia in heart failure (HF) patients. To determinate the responders patients, hemodynamic parameters are difficult to evaluate during follow-up, due to the invasivity of the procedures. We compare hemodynamic response to CRT with cardiac output, not invasively detected (CO2 rebreathing method, Innocor system), with conventional clinical, functional and echocardiographic parameters.

Methods: We enrolled 29 patients affected by end-stage dilated cardiomyopathy treated with CRT-P/CRT-D according to the latest guidelines (NYHA class II-IV, left ventricular ejection fraction [LVEF] ≤ 35%, QRS ≥ 120 ms, sinus rhythm, optimal medical therapy). Patients were evaluated before and after CRT (3 months), considering: NYHA class, Quality of Life score (Minnesota Living with Heart Failure questionnaire), QRS width, echocardiographic parameters (diastolic and systolic left ventricular volumes and related LVEF), six minutes walking test (6MWT) and cardiac output (detected with Innocor system).

Results: Our data showed a significant improvement in Innocor cardiac output 3 months after CRT implant compared to baseline (4.01±0.72 vs 4.48±0.59 l/min, p=0.001). The percentage improvement in cardiac output correlates with the percentage increase in LVEF (25±6% vs 30±7%; r=0.541). The correlation is not statistically significant with NYHA class (from 2.52±0.73 to 1.78±0.60; r=0.098), QoL (from 22.57±15.37 to 9.91±9.14 score; r=0.231) and exercise tolerance (from 390±50 to 437±54 meters; r=0.144).

Conclusions: The Innocor system is a promising non-invasive method to assess the cardiac output at baseline and during follow up in HF patients treated with CRT.

References

Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med 2002; 346: 1845-53. http://dx.doi.org/10.1056/NEJMoa013168 DOI: https://doi.org/10.1056/NEJMoa013168

Auricchio A, Stellbrink C, Sack S, Block M, Vogt J, Bakker P, et al. Long-term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol 2002; 39: 2026-2033. http://dx.doi.org/10.1016/S0735-1097(02)01895-8 DOI: https://doi.org/10.1016/S0735-1097(02)01895-8

Higgins SL, Hummel JD, Niazi IK, Giudici MC, Worley SJ, Saxon LA, et al. Cardiac resynchronization therapy for the treatment of heart failure in patients with intraventricular conduction delay and malignant ventricular tachyarrhythmias. J Am Coll Cardiol 2003; 42: 1454-1459. http://dx.doi.org/10.1016/S0735-1097(03)01042-8 DOI: https://doi.org/10.1016/S0735-1097(03)01042-8

Young JB, Abraham WT, Smith AL, Leon AR, Lieberman R, Wilkoff B, et al. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD trial. JAMA 2003; 289: 2685-2694. http://dx.doi.org/10.1001/jama.289.20.2685 DOI: https://doi.org/10.1001/jama.289.20.2685

Abraham WT, Young JB, Leon AR, Adler S, Bank AJ, Hall SA, et al. Effects of cardiac resynchronization on disease progression in patients with left ventricular systolic dysfunction, an indication for an implantable cardioverter-defibrillator, and mildly symptomatic chronic heart failure. Circulation 2004; 110: 2864-2868. http://dx.doi.org/10.1161/01.CIR.0000146336.92331.D1 DOI: https://doi.org/10.1161/01.CIR.0000146336.92331.D1

Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure (COMPANION). N Engl J Med 2004; 350: 2140-2150. http://dx.doi.org/10.1056/NEJMoa032423 DOI: https://doi.org/10.1056/NEJMoa032423

Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L, Tavazzi L. Longer-term effects of cardiac resynchronization therapy on mortality in heart failure (the CArdiac REsynchronization-Heart Failure [CARE-HF] trial extension phase). Eur Heart J 2006; 27: 1928-1932. http://dx.doi.org/10.1093/eurheartj/ehl099 DOI: https://doi.org/10.1093/eurheartj/ehl099

Moss AJ, Jackson Hall W, Cannom DS, Klein H, Brown MW, Daubert JP, et al. Cardiac-Resynchronization Therapy for the Prevention of Heart-Failure Events. N Engl J Med 2009; 361: 1329-1338. http://dx.doi.org/10.1056/NEJMoa0906431 DOI: https://doi.org/10.1056/NEJMoa0906431

St. John Sutton M, Ghio S, Plapperd T, Tavazzi L, Scelsi L, Daubert C, et al. Cardiac resynchronization induces major structural and functional reverse remodelling in patients with New York Heart Association class I/II heart failure. Circulation 2009; 120: 1858-1865. http://dx.doi.org/10.1161/CIRCULATIONAHA.108.818724 DOI: https://doi.org/10.1161/CIRCULATIONAHA.108.818724

Chung ES, Leon AR, Tavazzi L, Sun JP, Nihoyannopoulos P, Merlino J, et al. Results of the predictors of response to CRT (PROSPECT) trial. Circulation 2008; 117: 2608-2616. http://dx.doi.org/10.1161/CIRCULATIONAHA.107.743120 DOI: https://doi.org/10.1161/CIRCULATIONAHA.107.743120

Pohost GM, Hung L, Doyle M. Clinical use of cardiovascular magnetic resonance. Circulation 2003; 108: 647-653. http://dx.doi.org/10.1161/01.CIR.0000083233.86078.3E DOI: https://doi.org/10.1161/01.CIR.0000083233.86078.3E

Damgaard M, Norsk P. Effects of ventilation on cardiac output determined by inert gas rebreathing. Clin Physiol Funct Imaging 2005; 25: 142-147. http://dx.doi.org/10.1111/j.1475-097X.2005.00602.x DOI: https://doi.org/10.1111/j.1475-097X.2005.00602.x

Trinkmann F, Berger M, Hoffmann U, Borggrefe M, Kaden JJ, Saur J. A comparative evaluation of electrical velocimetry and inert gas rebreathing for the non-invasive assessment of cardiac output. Clin Res Cardiol 2011; 100: 935-943. http://dx.doi.org/10.1007/s00392-011-0329-9 DOI: https://doi.org/10.1007/s00392-011-0329-9

Rector TS, Kubo SH, Cohn JN. Patients’ self assessment of their congestive heart failure, part 2: content, reliability, and validity of a new measure, the Minnesota Living with Heart Failure questionnaire. Heart Failure 1987; 3: 198-209.

Dong L, Wang JA, Jiang CY. Validation of the use of foreign gas rebreathing method for non-invasive determination of cardiac output in heart disease patients. J Zhejiang Univ Sci B 2005; 6: 1157-1162. http://dx.doi.org/10.1631/jzus.2005.B1157 DOI: https://doi.org/10.1631/jzus.2005.B1157

Gabrielsen A, Videbaek R, Schou M, Damgaard M, Kastrup J, Norsk P. Non-invasive measurement of cardiac output in heart failure patients using a new foreign gas rebreathing technique. Clin Sci 2002; 102: 247-252. http://dx.doi.org/10.1042/cs1020247 DOI: https://doi.org/10.1042/cs1020247

Saur J, Fluechter S, Trinkmann F, Papavassiliu T, Schoenberg S, Weissmann J, et al. Noninvasive determination of cardiac output by the inert-gas rebreathing method. Comparison with cardiovascular magnetic resonance imaging Cardiology 2009; 114: 247-254. http://dx.doi.org/10.1159/000232407 DOI: https://doi.org/10.1159/000232407

Agostoni P, Cattadori G, Apostolo A, Contini M, Palermo P, Marenzi G, Wasserman K. Noninvasive measurement of cardiac output during exercise by inert gas rebreathing technique: a new tool for heart failure evaluation. J Am Coll Cardiol 2005; 46: 1779-1781. http://dx.doi.org/10.1016/j.jacc.2005.08.005 DOI: https://doi.org/10.1016/j.jacc.2005.08.005

Reutershan J, Kapp T, Unertl K, Fretschner R. Noninvasive determination of cardiac output in ventilated patients. Clinical evaluation of a simplified quick method. Anaesthesist 2003; 52: 778-786. http://dx.doi.org/10.1007/s00101-003-0547-2 DOI: https://doi.org/10.1007/s00101-003-0547-2

Christensen P, Clemensen P, Andersen PK, Henneberg SW. Thermodilution versus inert gas rebreathing for estimation of effective pulmonary blood flow. Crit Care Med 2000; 28: 51-56. http://dx.doi.org/10.1097/00003246-200001000-00008 DOI: https://doi.org/10.1097/00003246-200001000-00008

Downloads

Published

2016-08-16

How to Cite

Occhetta, E., Devecchi, C., Devecchi, P., Facchini, E., Di Ruocco, V., Plebani, L., & Nicola Marino, P. (2016). Non Invasive Cardiac Output Evaluation with CO2 Rebreathing Method for CRT Patients. International Journal of Statistics in Medical Research, 5(3), 175–181. https://doi.org/10.6000/1929-6029.2016.05.03.3

Issue

Section

Special Issue - Inference in Clinical Experiments